URGENT MEDTRONIC MEDICAL DEVICE RECALL

Recall of Medtronic Minimed Quick-SetŪ Infusion Sets Models MMT-396, MMT-397, MMT-398 and MMT-399 (Lots Starting with 8).

Recall of Medtronic Minimed Quick-Set® Infusion Sets Models MMT-396, MMT-397, MMT-398 and MMT-399 (Lots Starting with 8). Medtronic has notified us of their voluntary recall of Quick-Set® infusion sets that have lot numbers starting with the number "8". These infusion sets are used with the MiniMed Paradigm® Insulin pumps. They have asked us to assist in this recall action because they have identified a small percent of infusion sets that may not work properly.

The situation is related to the tubing connector, which may not allow the insulin pump to vent properly. Venting is necessary to equalize the pressure in the reservoir compartment with the surround atmosphere. If the vent does not work properly, this could potentially result in too much or too little insulin being delivered and may lead to serious injury or death.

Actions to take:
Step 1. Stop using "Lot 8" Quick-Set® infusion sets right away
Step 2. Contact us to replace your Quick-Set® infusion sets.

We will promptly ship replacement sets to you and provide pre-paid return shipping labels for you to return your recalled Medtronic Minimed Quick-Sets® to us.

To begin your replacement process, call (800)746-7505 or (972)243-2727